Blast Furnace Iron Block

CRM Category: Cast Irons
Material Form: 48 x 42 x 12mm Block
Commodity: Alloys
Type: Certified Reference Material
Product Code: SCRM 674/1

SCRM 674/1
£652.85
Accreditations

Chemical Content

ElementMass Fraction in %Uncertainty in Mass Fraction in %
Si0.000070.00006
Fe0.000070.00006
Cu0.000070.00006
Mn0.000070.00006
Mg0.000070.00006
Cr0.000070.00006
Ni0.000070.00006
Zn0.000070.00006
Ti0.000070.00006
Ag0.000070.00006

Properties

Density
6.99g/cm³

Do you have a specific question?

Instant access to experts who can answer your questions

Available 9am - 5pm Mon-Fri (UK time)

Product Description

The material for this SCRM was prepared by Replicast Ltd. at Sheffield, U.K using a special method of casting known to provide material of uniform composition in a form suitable for use as a Certified Reference Material (CRM) for optical emission spectrometric analysis. Blocks from this cast have been shown, by statistically designed procedures, to provide reproducible results using optical emission spectrometry. SCRM 674/1 is sold in the form of chill cast rectangular blocks, each approximately 48mm x 42mm x 12mm thick with a single chilled working face. Spectroscopic reproducibility has been shown to be reliable to a depth of 5mm below the original surface of this block. Sparking must be made on the fully ground surface only and the sample should be discarded when this face has been ground back as far as the small shoulder around the edge of the sample. The preparation of representative samples for chemical analysis and the certification by co-operative analysis was undertaken by Bureau of Analysed Samples Ltd. Bureau of Analysed Samples Ltd is a United Kingdom Accreditation Service (UKAS) Accredited Reference Material Producer, No 4004, and, as the Producer of SCRM 674/1 as defined in ISO Guide 34:2009 section 3.1, is fully responsible for assigning the certified values and their uncertainties in accordance with ISO Guides 34:2009 and 35:2006.

View certificate

Notes on methods used

CARBON
Analysts Nos. 1, 2, 4 and 5 determined carbon by high frequency combustion and infrared absorption. Analysts Nos. 3 and 6 determined carbon using non-aqueous titration according to the Standard Method BS 6200:3.8.2:1991.
SILICON
Analyst No. 1 determined silicon photometrically as silicophosphomolybdate without extraction. Nos. 2 and 6 determined silicon gravimetrically, according to BS 6200:3.26.1:1995. Nos. 3, 4 and 5 used inductively coupled plasma optical emission spectrometry (ICP-OES).
MANGANESE
Analysts Nos. 1, 2, 4 and 5 used ICP-OES. Nos. 3 and 6 determined manganese photometrically after oxidation with potassium periodate according to BS 6200:3.18.2:1995.
PHOSPHORUS
Analysts Nos. 1, 2, 4 and 5 used ICP-OES. Nos. 3 and 6 determined manganese photometrically after oxidation with potassium periodate according to BS 6200:3.18.2:1995
SULPHUR
Analyst Nos.1, 2, 4 and 5 determined sulphur using high frequency combustion and infrared absorption. Analyst No. 6 determined sulphur using oxidation/reduction titration after combustion.
CHROMIUM
Analysts Nos. 1, 2, 3, 4 and 5 determined chromium using ICP-OES. Analyst No.6 used flame atomic absorption spectrometry (FAAS).
MOLYBDENUM
Analysts Nos. 1, 2, 4 and 5 determined molybdenum using ICP-OES. Analyst No. 6 used FAAS
NICKEL
Analysts Nos. 1, 2, 3, 4 and 5 determined nickel using ICP-OES. Analyst No.6 used FAAS.
ALUMINIUM
Analysts Nos. 1, 2, 3, 4 and 5 determined aluminium using ICP-OES. Analyst No.6 used FAAS, according to BS6200: 3.1.4:1990.
ARSENIC
Analysts Nos. 1, 2, 4 and 5 determined arsenic using ICP-OES. Analyst No.6 determined As photometrically with silver diethyldithiocarbamate after separation as arsine
COBALT
All Analysts determined cobalt by ICP-OES
COPPER
Analysts Nos. 1, 2, 3, 4 and 5 determined copper using ICP-OES. Analyst No.6 used FAAS according to BS EN 24943:1990.
NITROGEN
Analysts Nos. 1, 2, 4 and 5 determined nitrogen using thermal conductivity. Analyst No.6 determined nitrogen titrimetrically after distillation as ammonia according to the Standard Method BS 6200:3.22.1:1992.
TIN
All analysts determined tin using ICP-OES.
TITANIUM
All Analysts determined titanium using ICP-OES.
VANADIUM
All Analysts determined vanadium by ICP-OES
NIOBIUM
All Analysts determined niobium using ICP-OES.
CALCIUM
Analyst No. 1 determined calcium using ICP-OES. Analysts Nos. 4 and 6 used FAAS.

Subscribe to our newsletter

Sign up to recieve the latest news and offers